202 Instructions for use Babyleo TN500 SW 1.0n
Annex
15.3 Automatic selftest
The operational readiness of the device is checked by means of a selftest. The
following table shows which device functions are automatically tested.
Test After
switching
on
Continu-
ously
Other
Power supply to the device X X
Consumer (radiant warmer,
convective heater, humidi-
fier heater)
XX
– Safety shut-down X Every 24 hours
Electronic components,
in areas flushed with
oxygen:
– Safety shut-down X
– Safety test X
Alarm loudspeaker:
– Acoustic and optical
tests by the user
X
2 acoustic signals sound;
the alarm light lights yel-
low and red
– Electrical test When alarms of high and
medium priority are initi-
ated
Sensor box X
Water valve For the 1st use
Fan impeller:
– Installed X
– Rotational speed X
Safety function of heated
mattress
XX
All sensors X X
Scale Automatic
Device height adjustment When used
Bed-tilt mechanism When used