Instructions for use Babyleo TN500 SW 1.0n 161
Troubleshooting
10.2 Power failure alarm
When the mains power supply is restored, the device continues the warming
therapy with the settings that were activated before the power supply failure.
!! 15 Weaning failed The next temperature step
for weaning could not be
reached.
Check the settings and adjust
them if applicable.
Check the patient's condition.
! 111 Weaning temperature not
reached
The next weaning tempera-
ture step could not be
reached. The previous set-
ting for air temperature is
applied.
Check the settings and adjust
them if applicable.
Check the patient's condition.
Alarm pri-
ority
Alarm Cause Remedy
Condition Cause Remedy
A loud acoustic signal
sounds. The failure is also
signaled on the display by
a red indicator light for
missing mains power sup-
ply. The user interface is
blank. The height of the
device cannot be adjusted
using the foot pedals. The
mattress tray control panel
cannot be operated.
The mains plug has been
disconnected without first
switching off the device
using the Start/standby
key on the display unit.
Switch off the device at the
main switch.
Before disconnecting the
mains plug, switch off the
device at the main switch.
For further information,
see the following chapter:
''Switching off the
device'' (page 67)
The mains power supply
has failed.
Restore the mains power
supply.