Instructions for use Babyleo TN500 SW 1.0n 155
Troubleshooting
! 169 Check oxygen supply The oxygen hose is not con-
nected.
Make sure that the oxygen
hose is correctly connected.
The oxygen cylinder is
empty.
Check the oxygen cylinder.
The valve of the oxygen cyl-
inder is closed.
Make sure that the valve of
the oxygen cylinder is open.
The oxygen supply pressure
is too high.
Check the oxygen supply, the
central supply, and the pres-
sure reducer.
An internal fault occurred. Contact specialized service
personnel.
! 219 Check patient's condition The radiant warmer has
been running at more than
30 % for 14 minutes.
Check the set value.
To continue with the last
active settings for the radiant
warmer, press the "Alarm
reset" key.
!! 53 Check settings New settings are applied
automatically.
Check the patient's condition.
Check the settings and adjust
them if applicable.
!! 52 Check settings Settings were imported from
a USB mass storage device.
Check the settings and adjust
them if applicable.
!! 51 Check settings An internal fault occurred. Check the settings and adjust
them if applicable.
! 108 Data export failed Data export failed. Repeat the data export.
The USB mass storage
device is faulty.
Use another USB mass stor-
age device.
Contact specialized service
personnel.
! 112 Data import failed Data import failed. Repeat the data import.
Check the USB mass storage
device.
Contact specialized service
personnel.
! 157 Data loss An internal fault occurred
during data storage.
Check the current settings
and the default settings.
Contact specialized service
personnel.
!! 69 Device failure, heating off An internal fault occurred. Put the device out of opera-
tion.
Contact specialized service
personnel.
!! 68 Fan faulty The fan impeller is not
installed.
Switch off the device and
check the fan impeller.
The fan is faulty. Contact specialized service
personnel.
Alarm pri-
ority
Alarm Cause Remedy