52 Instructions for use Babyleo TN500 SW 1.0n
Operating concept
4.4.2 Opening and closing the hood
The hood can be opened and closed manually.
CAUTION
Risk of injury due to unexpected opening or closing of the hood
Opening or closing the hood unexpectedly may cause head or other injuries to
people nearby.
► Make sure that no-one else is using the device at the same time and that there is
no-one else in the vicinity of the hood.
CAUTION
Risk from falling objects
Objects placed on the hood may fall off, injuring the patient.
► Remove all objects before opening the hood.
Procedure:
1 To open the hood, take hold of the handle at the side of the hood and lift it
forwards and up.
The hood moves up.
2 To close the hood, pull the handle at the side of the hood downwards.
The device closes the hood automatically at the end.
4.5 Hand ports and access panels
This chapter describes how to operate the hand ports and access panels.
4.5.1 Overview of the access options
The patient can be cared for in the patient compartment via the following panels
and ports:
– On each side of the device there is a side access panel (1) with 2 hand ports (4).
– On the front of the device there is the front access panel (2) with one hand port
(3).
Fig. 25 Access panels and hand ports
When the hand ports or access panels are open, a warm air curtain remains in
place, preventing the patient compartment from cooling down.