Instructions for use Babyleo TN500 SW 1.0n 63
Operation
6 Operation
6.1 Checking for operational readiness
This chapter describes how to check the device for operational readiness. For
further information, see the following chapter: ''Automatic selftest'' (page 202)
Prerequisites:
The device has been acclimatized in the environment of use for 24 hours prior to
use.
The device has been reprocessed and assembled ready for operation.
The mains plug is plugged in.
WARNING
Risk to patient due to undetected damage to the device
Reprocessing, wear and tear, and improper storage may damage parts of the
device such that they no longer function correctly.
Comply with the ambient conditions for operation and storage of the device.
Before operating the device, always carry out the test steps listed in the
"Checking for operational readiness" chapter.
Only operate the device when all test steps and the automatic selftest have been
completed successfully.
Position the device so that the mains plug is readily accessible at all times. The
device can only be safely disconnected from the mains power supply by
unplugging the mains plug.
Procedure:
Always perform the following test steps before operating the device:
1 Check that all parts of the device are free from cracks and sharp edges,
especially the hood, the access panels, and the mattress tray.
2 Check that all cables and hoses are free from damage and are routed correctly.
3 Check that the heated mattress (option) is free from damage.
4 Check that all access panels and hand ports are free from damage and close
securely.
5 Check that 2 acoustic signals sound and the alarm light shows yellow and red
when the device is switched on.
6 After switching the device on, check that the touch screen and rotary knob are
working correctly.
7 Check whether the O
2 sensors need to be adjusted. For further information, see
the following chapter: ''Adjusting the O
2 sensors'' (page 99)
Only operate the device when all test steps and the automatic selftest have been
completed successfully.