192 Instructions for use Babyleo TN500 SW 1.0n
Technical data
14.5 Materials
Bisphenol A (BPA) The patient compartment of the device (without
accessories) is free from bisphenol A
Latex The device (without accessories) contains no
latex-containing components
14.6 Data exchange and interfaces
Port for central alarm Output for the connection to intercommunica-
tion system (nurse call)
Only connect devices that fulfill the require-
ments of the standard IEC 60950-1 on
unearthed SELV circuits or the requirements of
the standard IEC 60601-1 on accessible sec-
ondary circuits with max. 60 V DC nominal volt-
age.
Type ODU MEDI-SNAP 3-pole with 205° coding (J)
Operating voltage 30 V max.
Current 250 mA max.
Power 7.5 W max.
Floating changeover contact
1 - Common port
2 - Alarm
3 - No alarm
4 - Nurse call
COM port Only connect devices that fulfill the require-
ments of the standard IEC 60950-1 on
unearthed SELV circuits or the requirements of
the standard IEC 60601-1 on accessible sec-
ondary circuits with max. 60 V DC nominal volt-
age.
Type 9-pole D-sub, electrically isolated:
1.5 kV against internal electronics
1.0 kV against housing
Protocol MEDIBUS.X
Alarm delay Typically < 2 s
Baud rate 1200, 2400, 4800, 9600, 19200, 38400 baud