Instructions for use Babyleo TN500 SW 1.0n 191
Technical data
Setting range 34 to 38 °C (93.2 to 100.4 °F)
>37 °C (98.6 °F), after confirmation
Heated mattress (option)
Temperature measurement (mattress core) NTC, 4 x
Setting range Off, 35 to 39 °C (95 to 102 °F), AUTO
Oxygen regulation
Measurement principle Electrochemical, with automatic barometric
pressure compensation
Measurement range Normal range: 18 to 65 Vol%
Extended range: 66 to 99 Vol%
Measurement uncertainty in normal range ± (2.5 Vol% + 2.5 % of measured value)
Influencing factors Relative ambient humidity <1.5 %
Setting range Off, 21 to 65 Vol%
>40 Vol%, after confirmation
O
2 rise time from 21 to 65 Vol% <10 minutes
Humidifier regulation
Measurement principle Capacitive
Measurement range Normal range: 30 to 99 % R H
Extended range: 10 to 29 % R H
Measurement uncertainty in normal range ±10 %
Setting range Off, 30 to 99 % r. H. in steps of 1 %, AUTO
Noise measurement in patient compartment
Measurement range 40 to 90 dB(A)
Light measurement in patient compartment
Measurement range 3 to 999 Lux (1 to 93 fc)
Scale
The measuring accuracy depends on the local
geographic conditions. The geographical loca-
tion where the device is operated is stored in
the device ex works. The set parameters are
listed in the service dialog and on a product
label on the rear of the device.
Measuring range (incl. tare) 200 to 10000 g (0.44 lb to 22 lb)
Display range 0 to 10045 g (0 to 22 lb 2 oz)
Measurement uncertainty 0 to 2500 g (0 to 5.5 lb): 2.5 g (0.0055 lb)
for loads over 2500 g (5.5 to 22 lb): 5 g
(0.011 lb)
Resolution (OIML version) 5 g (0.011 lb)
Resolution (standard version) 1 g (1 oz)
14.4 Measured value display (continued)