Instructions for use Babyleo TN500 SW 1.0n 57
Assembly and preparation
Additional information:
The following loading limits apply if only 1 pivot arm is fitted to the device:
Duct rail on the side of the hood arm:
Duct rail on the side without the hood arm:
Or:
The following loading limits apply if a total of 2 pivot arms are fitted to the device:
Either 15 kg (33.1 lbs) or 12 kg (26.5 lbs) can be fitted to each duct rail.
The user can decide which weight is fitted to which side of the device.
5.2 Nurse call
5.2.1 Information on the nurse call
The nurse call is used for transmitting high- and medium-priority alarms to a central
hospital alarm system.For further information, see the following chapter: ''Alarm –
Cause – Remedy'' (page 153)
If the acoustic alarm signal of the device fails or the power supply is interrupted, the
nurse call is still activated.
If the "Audio paused" key is pressed during an alarm, the acoustic alarm signal on
the device and the nurse call are suppressed.During this time, any new alarms that
occur do not trigger a nurse call.
8 X-ray tray 3 kg (6.6 lbs) X-ray cassette or X-ray
detector
9 Mattress 5 kg (11 lbs) Patient and patient acces-
sories
Maximum load Arm length
15 kg (33.1 lbs) max. 300 mm (11.81 in)
Maximum load Arm length
20 kg (44.1 lbs) max. 300 mm (11.81 in)
Maximum load Arm length
25 kg (55.1 lbs) max. 250 mm (9.84 in)