156 Instructions for use Babyleo TN500 SW 1.0n
Troubleshooting
!! 67 Fan motor faulty, heating
off
The fan impeller is faulty. Switch off the device.
Contact specialized service
personnel.
! 158 Foot pedal faulty The foot pedal has been
pressed for more than one
minute.
Release the foot pedal.
Contact specialized service
personnel.
! 167 Hood not closed, warmer
off
The position of the hood can-
not be determined.
Close the hood properly.
Contact specialized service
personnel.
! 166 Hood position unclear,
warmer off
The position of the hood can-
not be determined.
Close the hood. Only incuba-
tor operation is available.
Contact specialized service
personnel.
! 216 Humidifier faulty An internal fault occurred. Set the therapy control for
humidity to "Off".
Contact specialized service
personnel.
! 218 Humidity low The hood, the side access
panels, or the hand ports are
open.
Close the hood, the side
access panels, or the hand
ports.
The water level of the humid-
ification system is low.
Make sure that the water
supply is correctly con-
nected.
Refill the water supply.
The humidity sensor is faulty. Contact specialized service
personnel.
!! 58 Humidity not confirmed The setting was changed,
but the change was not con-
firmed.
Adjust the setting for humidity
and press the rotary knob to
confirm.
!! 66 Internal temperature high An internal fault occurred. Switch off the device.
Contact specialized service
personnel.
!! 100 Kangaroo mode not con-
firmed
The mode was selected, but
it was not confirmed.
Select kangaroo mode and
press the rotary knob to con-
firm.
! 159 Lifting column faulty An internal fault occurred. Contact specialized service
personnel.
! 105 Light sensor faulty An internal fault occurred. Contact specialized service
personnel.
! 160 Loudspeaker faulty An internal fault of the pri-
mary acoustic alarm system
occurred.
Contact specialized service
personnel.
! 156 Maintenance due The set maintenance inter-
val has been exceeded.
Contact specialized service
personnel.
!! 61 Manual mode not con-
firmed
The mode was selected, but
it was not confirmed.
Select manual mode and
press the rotary knob to con-
firm.
Alarm pri-
ority
Alarm Cause Remedy