96 Instructions for use Babyleo TN500 SW 1.0n
Operation
Procedure:
1 Touch the Humidity therapy control (1) in the therapy bar.
The therapy control turns yellow.
Fig. 50 Therapy control for humidity
2 Set the value for the humidity manually using the rotary knob.
or
3 Turn the rotary knob to the right as far as it will go and set the Auto function.
The device sets the humidity automatically according to the air temperature.
4 Press the rotary knob to confirm.
The therapy control turns dark green.
The value for the humidity is set.
6.12 Oxygen (option)
This chapter describes how to enrich the air in the patient compartment with oxygen
(option).
6.12.1 Overview of the oxygen supply
The air in the patient compartment can be enriched with oxygen. The oxygen
concentration in the patient compartment is measured by O
2 sensors.
The table below shows the setting ranges for the supply of oxygen:
To set the oxygen supply in the extended range, first confirm the extended range
by pressing the rotary knob.
27240
1
Setting range Extended range
Off, 21 to 40 % 41 to 65 %