48 Instructions for use Babyleo TN500 SW 1.0n
Operating concept
4.2 Adjusting the device height
The height of the device can be adjusted with the lifting column. The lifting column
is controlled by means of 2 foot pedals on the trolley.
Prerequisites:
The device is in operation.
The device height adjustment mechanism is not locked.
WARNING
Risk of electric shock
If height adjustment is actuated and the power cable is trapped or wrapped around
a castor, the power cable may tear.
When positioning the device, ensure that the power cable is not trapped.
Be aware of the length of the power cable when operating the height
adjustment.
During intrahospital transport, do not push the device over the power cable.
WARNING
Risk of injury by collision with compressed gas cylinders
When height adjustment is actuated, accessories on the duct rail may collide with
the gas cylinders.
Actuate height adjustment with caution and stop it promptly.
WARNING
Risk to patient if hoses and cables are too short
Hoses and cables may be subjected to tensile loading when opening the hood,
moving the device, or actuating the device height adjustment. This risk applies
especially if the hoses and cables are routed into the patient compartment across
the hood.
Choose hoses and cables that are long enough to perform the above
movements safely.
Do not route hoses and cables across the hood; always pass them into the
patient compartment via the hose grommets.
Pay attention to the hoses and cables when carrying out the above movements.
WARNING
Unintentional disconnection of hose and cable connections
Possible interruption to patient treatment due to uncontrolled movement of the
device via height adjustment or the trolley.
When using the height adjustment mechanism and the trolley, always ensure
that no hoses and cables become disconnected.
Lock the double castors on the trolley with the locking brakes.