Instructions for use Babyleo TN500 SW 1.0n 85
Operation
6.8.8 Ending skin temperature mode
As soon as a different mode is selected in the therapy bar, skin temperature mode
is ended.
Procedure:
1 Touch a different tab in the therapy bar.
The selected tab turns yellow.
2 Press the rotary knob to confirm.
The selected tab turns dark green.
✓ Skin temperature mode is ended.
6.9 Air temperature mode
This chapter describes air temperature mode. Air temperature mode can be set in
incubator operation.
6.9.1 Overview of air temperature mode
In air temperature mode the user determines the appropriate temperature in the
patient compartment. The temperature is measured by a sensor in the patient
compartment and compared with the set target value. The air temperature can be
set in steps of 0.1 °C (0.1 °F). The following table shows the temperature ranges:
To set the temperature in the extended range, first confirm the extended range
by pressing the rotary knob.
The target air temperature must be at least 3 °C (5.4 °F) above the ambient
temperature in order for the device to reach the desired temperature.
In air temperature mode the patient's skin temperature can be monitored with skin
temperature sensors.
Extended range Setting range Extended range
20 to 27.9 °C
(68 to 82.3 °F)
28 to 37 °C
(82.4 to 98.6 °F)
37.1 to 39 °C
(98.7 to 102.2 °F)