142 Instructions for use Babyleo TN500 SW 1.0n
Configuration
8.4 System setup
8.4.1 Specifying the device settings
The following tables show the settings options in the System setup dialog window.
The description assumes the maximum device configuration. The actual device
may differ in its functions from the descriptions below.
Various settings are possible for the device:
Factory settings
Dräger delivers the device with factory settings that are used when starting the
device for the first time. Users can reset the device to factory settings.
User-specific settings
The user can define specific settings for the device. The settings (e.g., therapy
values, alarm limits, or screen layout) can be made in the System setup dialog
window. User-specific settings are saved automatically once they are confirmed.
User-specific settings can be transferred to other devices via the Import/export
function. For further information, see the following chapter: ''Transferring data to
another device'' (page 134)
User-specific settings are automatically applied in the following situations:
When the device is restarted after it has been switched off for more than 2
hours.
If the New patient button was activated when the device was switched on.
Patient-specific settings
The user can adjust the factory settings or the user-specific settings to suit the
current patient. The values for the patient-specific settings are set on the main
screen via the therapy bar (modes and therapy controls). Patient-specific alarms
are set via the main screen in the Alarms dialog window. The patient-specific
settings also include the information in System setup > Patient dialog window.
Patient-specific settings can be transferred to other devices via the Import/export
function. For further information, see the following chapter: ''Transferring data to
another device'' (page 134)
Patient-specific settings are automatically set in the following situations:
When the device is restarted after it has been switched off for less than 2 hours.
If the Current patient button was activated when the device was switched on.
The configuration password must be entered for some settings. For further
information, see the following section: ''Configuration password for Babyleo
TN500 Software 1.0n'' (page 213).
8.4.1.1 Resetting to factory settings
Certain tabs in the System setup dialog window have a button for resetting the
user-specific settings to the factory settings.
1 Open the relevant tab, e.g., Screen layout, Alarms, Therapy, System.