Instructions for use Babyleo TN500 SW 1.0n 141
Configuration
8.3.3 Measured values and settings
All current measured values and current settings for the thermoregulation are
displayed in the Values/settings tab.
8.3.4 Logbook
All events and alarms are displayed in chronological order in the Logbook tab. The
latest event/latest alarm is at the top. Earlier events and alarms can be displayed
using the arrow keys.
The entries in the logbook cannot be deleted and are preserved even after the
device is switched off and on or after a power supply failure. There is no record kept
of how long the device and alarm system were shut down. When the capacity of the
logbook is reached, the oldest entries are overwritten.
The following events and alarms are displayed:
– All alarms, their confirmation, and muting by the user
– Alarm limit changes
– Operating mode changes
– Therapy control changes
– Start and end of applications
– Events such as hood open, side access panel open
– Data import and export to USB mass storage device
– Device time changes
Userdefined Temperature Central skin temperature X
Peripheral skin tempera-
ture
X
Skin temperature differ-
ence
X
Air temperature X
Mattress core tempera-
ture
–
Patient com-
partment
Humidity –
Oxygen –
Heating Radiant warmer X
Convection heater X
Noise/light Noise level –
Light level –
Vertical tab Heading Parameter Selected in fac-
tory settings