Instructions for use Babyleo TN500 SW 1.0n 201
Annex
15.1.1 Safety mode in incubator operation and radiant warmer
operation
If a skin temperature regulation fault occurs in skin temperature mode, the device
generates an alarm and switches into a safety mode (fall-back). If one of the
following alarm messages is displayed, the device deactivates the heat supply:
15.2 Further information on heated mattress (option)
This chapter describes the special features of the heated mattress (option) when
used in incubator operation and radiant warmer operation.
15.2.1 Incubator operation and radiant warmer operation
The heated mattress can be used in incubator operation and in radiant warmer
operation:
Incubator operation
In incubator operation, the displayed mattress core temperature may be higher than
the temperature in the patient compartment.
Radiant warmer operation
In radiant warmer operation, the surface temperature of the mattress tray may be
higher than the displayed mattress core temperature due to the influence of the
radiant warmer.
Alarm Cause Effect
Air temperature too
high
Air temperature > 40 °C
(> 104 °F)
The convective heater is
switched off.
Centr. sensor miss-
ing, heating off
The central skin tem-
perature sensor (yellow)
is not connected.
The convective heater or the
radiant warmer is switched
off.
Centr. temp. sensor
faulty, heating off
The central skin tem-
perature sensor (yellow)
is faulty.
The convective heater or the
radiant warmer is switched
off.
Skin temperature high The tolerated deviation
for central skin tem-
perature (yellow) has
been exceeded. The
current central skin
temperature is too high.
The convective heater or the
radiant warmer is switched
off.