Instructions for use Babyleo TN500 SW 1.0n 19
Safety-related information
2.13.4 Air flow in the warm air duct
If the warm air duct in the device becomes blocked or covered, thermoregulation is
disrupted. The hot air flow may reach the patient directly, causing burns.
► Do not block or cover the warm air duct (1) and do not deflect the hot air flow to
the patient using objects (e.g., blankets).
The warm air duct passes through the air ducts of the X-ray flap. The convective
heater directs the warm air through the air ducts of the X-ray flap into the patient
compartment.
Fig. 3 Air ducts of the X-ray flap
2.14 Therapy and applications
2.14.1 Ambient conditions
If the permissible ambient conditions are not complied with, the values set for
humidity and temperature in the patient compartment may not be achieved. The
patient may be put at risk.
► Be aware of and comply with the permissible ambient conditions. For further
information, see the following chapter: ''Technical data'' (page 188)
2.14.2 Therapy settings
If the therapy settings are unsuitable for the patient, the patient may be put at risk.
► Before starting therapy, check the therapy settings carefully. Adjust the therapy
settings to the needs of the patient, e.g., after transferring patient data to a new
device, after transferring data via a USB port, after changing the operating
mode, when using the default settings.
► Measure the body core temperature at regular intervals using a separate
thermometer.
30231