Instructions for use Babyleo TN500 SW 1.0n 67
Operation
6.3 Switching off the device
This chapter describes how to switch off the device. Before switching off the device,
lower the lifting column to transfer height. This prevents the device from tipping over
or being damaged during a transfer. For further information, see the following
chapter: ''Intrahospital patient transfer with switched off warming
therapy'' (page 150).
Prerequisite:
– The device is in operation.
Procedure:
1 Adjust the height of the device using the foot pedals. The correct transfer height
is reached when the red hatched label on the lifting column is completely
covered.
2 Switch off the device on the display unit using the Start/standby key .
The Shut down dialog window is displayed.
3 Touch the Shut down button and confirm.
The device functions are shut down.
4 Switch off the device using the main switch on the rear of the main column
(position ).
Before disconnecting the mains plug, the device must be switched off at the
main switch. If the device is only switched off using the Start/standby key on
the display unit and the mains plug is then disconnected, an alarm signal will
sound after a few seconds.
5 Disconnect the mains plug.
6 Fasten the power cable to the device, e.g., by wrapping it around the standard
rail.