76 Instructions for use Babyleo TN500 SW 1.0n
Operation
6.7.2 Configuring the transition
This chapter describes how to configure the transition.
Prerequisite:
– The device is in operation.
WARNING
Risk to patient due to incorrect temperature settings
When switching between incubator operation and radiant warmer operation, the
temperature setting may subsequently be too high or too low.
► After opening or closing the hood, always check the temperature settings.
► After the transition, adjust the temperature settings to the patient's needs.
Procedure:
1 Touch the System setup... button in the main menu bar.
2 Touch the Therapy tab.
3 Enter the password and confirm with OK.
4 Select a button in the Thermoregulation tab:
✓ The transition is configured.
Button Meaning
Standard – In incubator operation the radiant warmer operates contin-
uously at low power.
– In radiant warmer operation the convection heater oper-
ates continuously at low power.
This ensures optimum warming during and after the tran-
sition.
Eco – In incubator operation the radiant warmer is switched off.
– In radiant warmer operation the convection heater is
switched off.
This saves energy.