Instructions for use Babyleo TN500 SW 1.0n 21
Safety-related information
2.14.6 Heated mattress
If the heated mattress (option) is used as the only source of heat, the patient may
be put at risk.
Provide the patient with the heated mattress and another source of heat,
especially if the patient is cold.
Monitor the body core temperature of the patient with a thermometer.
Do not leave the patient unattended.
If the heated mattress is connected, do not use radio-frequency surgical
equipment, endocardial catheters, or defibrillators.
2.14.7 Transdermal plasters
If a plaster treated with an active agent (transdermal plaster) is applied to the
patient's skin during warming therapy, too much of the active agent may be
released. This may put the patient at risk.
Do not apply transdermal plasters to the patient's skin during and prior to
warming therapy, or be aware of the possibility that the active agent may be
absorbed too rapidly.
2.14.8 Sensor unit
If the slots on the sensor unit become blocked, malfunction may be caused.
Make sure that the slots on the sensor unit are kept clear at all times.