82 Instructions for use Babyleo TN500 SW 1.0n
Operation
6.8.4 Starting skin temperature mode
WARNING
Risk of hypothermia or hyperthermia
If skin temperature mode is used for patients with a fever or in a state of shock, the
patient may become too hot or too cold.
Do not use skin temperature mode in cases of fever or shock.
Procedure:
1 Touch the Skin mode tab (1) in the therapy bar.
The tab turns yellow.
2 Press the rotary knob to confirm.
The tab turns dark green.
Fig. 41 Starting skin temperature mode
3 Touch the Skin temp. therapy control (2).
The therapy control turns yellow.
4 Set the target skin temperature using the rotary knob.
5 Press the rotary knob to confirm.
The therapy control turns dark green and displays the target value.
When the hood is closed, the measured central skin temperature determines
the air temperature in the patient compartment; when the hood is open, the
measured central skin temperature determines the radiant warmer power.
6 At regular intervals check that the skin temperature sensors are correctly
attached to the patient.
7 Check the skin temperature value at regular intervals and adjust it if necessary.
The patient is warmed in accordance with his or her needs.
The temperature is continuously monitored by skin temperature sensors.
8 Check the alarm limits for skin temperature and adjust them if necessary.