Problem solving
82
Instructions for use D-Vapor/D-Vapor 3000
Anesthesia workstations
equipped with a vaporizer de-
tection system display anesthet-
ic agent which is different from
the vaporizer.
Coding for plug-in adapter or va-
porizer is damaged, faulty, or in-
correctly fitted.
Check coding, have it re-in-
stalled if required; repair by spe-
cialized service personnel.
Monitor displays different anes-
thetic agent than indicated on
the vaporizer.
A different anesthetic agent has
just been used. High concentra-
tions of it are still present in the
breathing system.
Flush breathing system or wait
for gas to change.
Incorrect anesthetic agent or an-
esthetic agent mixture in the va-
porizer.
Check the vaporizer, drain and
blow off, see page 69 and 71.
Have the vaporizer repaired by
specialized service personnel.
Monitor has not yet switched
over after a change of anesthet-
ic agent.
Wait until the monitor switches
over.
Monitor defective. Replace monitor.
Control dial cannot be set to
concentration.
0 button not pressed. Press the 0 button.
Interlock not switched over.
Interlock jamming or another va-
porizer still switched on.
Switch off other vaporizer and
switch over Interlock.
Checks, see page 54 and 60.
Repair by specialized service
personnel.
The control dial can be moved
from the positions 0 and T with-
out the button being pressed.
(12 % mark can be exceeded
without pressing the button.)
0 button defective. Repair by specialized service
personnel.
Fault Cause Remedy