Instructions for use | Oxylog VE300 SW 1.n 105
Reprocessing
12.8 Assembly
– Reassemble the device using the information in section "Assembly and
preparation" on page 33 as a guide.
– Connect to the power supply and gas supply using the information in section
"Assembly and preparation" on page 33 as a guide.
– Check readiness for operation using the information in section "Getting started"
on page 47 as a guide.
Accessories
bag
Yes No No No If soiled
Carrying strap Yes No No No If soiled
Components Surface
disinfec-
tion with
cleaning
Manual
cleaning
followed by
disinfec-
tion by
immersion
Machine
cleaning
with ther-
mal disin-
fection
Steam
steriliza-
tion
Recom-
mended
reprocess-
ing intervals