7.6 O2/air mix
The inspiratory oxygen concentration FiO2 can be set to O2/air mix independent of
the ventilation mode. With O
2/air mix, ambient air is drawn in by the injector
principle of the device to achieve an FiO
2 concentration of approx. 40 %.
However, the O
2 concentration which can be realized depends on the mean airway
pressure and the inspiratory flow. The O
2 concentration can never be below 40 %.
This is demonstrated in the following schematic diagrams: