Instructions for use | Oxylog VE300 SW 1.n 41
Assembly and preparation
3. Connect the oxygen compressed gas hose to the device's quick-coupling port
(B) until it locks in place and the oxygen supply is assured.
4. Slowly close the valve of the built-in oxygen cylinder.
5.7 Connecting the breathing circuit
Sterilize or disinfect all reusable parts before every use. For further information,
see the following chapter: "Reprocessing", page 98.
CAUTION
Risk of patient injury
A bent or twisted breathing hose may put the patient at risk.
Make sure that the breathing hoses are not bent or twisted.
CAUTION
Risk of patient injury
A breathing hose that does not correspond to the device settings may put the
patient at risk.
When connecting a breathing hose, make sure that the hose settings in the
configuration dialog correspond to the connected hose. For further information,
see the following chapter: "Setting the breathing circuit", page 75.
5.7.1 Connecting the reusable breathing circuit for adults
Assembly of the breathing valve
WARNING
Risk of CO2 rebreathing
An incorrectly assembled or malfunctioning breathing valve may put the patient at
risk.
Do not remove, damage, or bend the rubber disk (A) in the housing.
30064
A

Table of Contents

Other manuals for Dräger Oxylog VE300

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Dräger Oxylog VE300 and is the answer not in the manual?

Dräger Oxylog VE300 Specifications

General IconGeneral
BrandDräger
ModelOxylog VE300
CategoryMedical Equipment
LanguageEnglish

Related product manuals