40 Instructions for use | Oxylog VE300 SW 1.n
Assembly and preparation
1. Connect the pressure reducer (A) to the oxygen cylinder (D).
2. Connect the oxygen compressed gas hose (B) to the pressure reducer (A).
3. Connect the oxygen compressed gas hose (B) to the device.
4. Hand-tighten the screw fitting (C) on the device.
5. Rotate the cylinder valve slowly and open it fully.
Connecting the device to a central gas supply by the quick-coupling port
To maintain continuous oxygen supply when switching from the built-in oxygen
cylinder supply to central gas supply, the device can also be connected to a central
gas supply system via the quick-coupling port while the cylinder supply is active.
1. Make sure that the patient is receiving oxygen from the built-in oxygen cylinder.
2. Connect the gas probe of the oxygen compressed gas hose (A) to the oxygen
terminal unit.
3024430248
A
B
D
C
A
B