76 Instructions for use | Oxylog VE300 SW 1.n
Configuration
9.6 Setting the CO2 cuvette type (option)
The device can be set for the use of disposable or reusable CO2 cuvettes.
To set the CO
2 cuvette type:
Touch the CO
2 cuvette button, set the desired cuvette type (Disposable or
Reusable), and confirm the setting.
9.7 Using the Bluetooth function (option)
Activating Bluetooth
The device can be connected with another suitable receiver via Bluetooth.
To activate the Bluetooth function:
Touch the Bluetooth Off button, set to Bluetooth On and confirm the setting.
The Bluetooth PIN for connecting with another receiver is displayed below the
ventilation category for 15 seconds.
Then, the Bluetooth symbol appears below the ventilation category to indicate that
the function is active.
Connecting an external device
Operating systems that use Bluetooth may not all be compatible with one
another.
1. Activate the Bluetooth function on both devices.
2. Start the connection from an external device (coupling).
3. In the external device, enter the PIN displayed in the notification field of the
ventilator. The devices are then coupled automatically.
After the devices are coupled, the ventilator automatically sends a patient log file in
CSV format.
This patient log file contains the following information:
–Events
–Settings
Measured values for the current ventilation (device is ventilating) or
Measured values for the last ventilation (device is in standby mode)
Alternatively, after coupling, measured values, waveforms, alarms, and settings can
be transferred in real time using the MEDIBUS.X protocol.
MEDIBUS.X is a software protocol for the transfer of data between Oxylog VE300
and an external medical or non-medical device (e.g., patient monitors or computers
for data management systems).