Instructions for use | Oxylog VE300 SW 1.n 121
Technical data
15.11 EMC Declaration
15.11.1 General information
This device was tested for electromagnetic compatibility using accessories from the
list of accessories.
WARNING
Risk of patient injury and device malfunction
The use of accessories which are not included in the list of accessories as well as
converters and cables which are not sold by Dräger as spare parts may increase
electromagnetic emissions and compromise the immunity of the device.
► Only use the accessories, converters, and cables indicated by the manufacturer.
► Other accessories may only be used if they do not compromise the
electromagnetic compatibility.
WARNING
Risk of patient injury and device malfunction
If the device is used in the direct vicinity of other devices, patients may be harmed
or the device may be damaged.
► Only use the device in the direct vicinity of other devices if Dräger has approved
this device arrangement.
► If no approval has been given by Dräger, ensure that this device functions
correctly in the desired arrangement before use.
► Observe the other devices’ instructions for use.
15.11.2 Electromagnetic environment
This device may only be used in environments specified in section "Environment of
use" on page 8.
Emissions Compliance
Radiated emissions Class B, group 1 (30 MHz to 1 GHz)
Conducted emissions Class B, group 1 (150 kHz to 30 MHz)
Immunity against Test level and required electromag-
netic environment
Electrostatic discharge (ESD)
(IEC 61000-4-2)
Contact discharge: ±8 kV
Air discharge: ±15 kV
Electrical fast transients (bursts)
(IEC 61000-4-4)
Power cable: ±2 kV
Longer signal input lines/output lines:
±1 kV
Surge on AC mains lines
(IEC 61000-4-5)
Phase conductor voltage – phase con-
ductor: ±1 kV
Phase conductor voltage – protective
ground conductor: ±2 kV
Magnetic fields at mains frequency
(IEC 61000-4-8)
50 Hz: 30 A/m