106 Instructions for use | Oxylog VE300 SW 1.n
Service
13 Service
13.1 Safety information
WARNING
Risk due to inappropriately reprocessed products
The product may be contaminated with infectious agents.
► Before service is performed and before the product is sent back for repair,
reprocess the product in accordance with the chapter "Reprocessing".
WARNING
Risk if service is not performed regularly
Wear and material fatigue on the components may lead to device failure and
malfunctions.
► Perform service activities at the specified intervals.
WARNING
Risk if service is not performed properly
Personal injury and property damage may occur if service is not performed
properly.
► Service must be performed by those target groups that are assigned to the
particular measure.
WARNING
Risk when the housing is being opened
Under the housing there are live electrical components, which may cause an
electric shock.
► The housing may only be opened by those target groups that are assigned to
that particular measure.
13.2 Definition of service terminology
13.3 Inspection
Designation applies to the Federal Republic of Germany; corresponds to the
“Recurring safety inspection” in the Federal Republic of Austria.
Concept Definition
Service All measures (inspection, maintenance, repair) intended to
maintain or restore the functional integrity of a product
Inspection Measures intended to determine and assess the current state
of a product
Maintenance Regular specified measures intended to maintain the functional
integrity of a product
Repair Measures intended to restore the functional integrity of a pro-
duct after a failure
Measure Interval Target group
Inspection and safety check
(Oxylog VE300 and CO
2 sensor)
Every 2 years Service personnel