102 Instructions for use | Oxylog VE300 SW 1.n
Reprocessing
12.4.2 Classification of device-specific components
The entire device must be reprocessed. Observe the following classification and the
instructions for use for the components.
12.5 Recommendations for reprocessing
12.5.1 Validated reprocessing procedure
The effectiveness of the reprocessing procedures listed has been validated by
independent laboratories that are certified to the standard ISO 17025.
At the time of validation, the following reprocessing procedures showed good
material compatibility and effectiveness:
Manual cleaning and disinfection
Machine cleaning and thermal disinfection
Surface disinfection
Classification Device-specific components
Non-critical The entire device and all components that can be disas-
sembled (e.g., the accessories bag, CO
2 sensor,
O
2 hose)
Semi-critical Breathing valve, flow sensor, pressure sensor, breathing
hose with measuring lines, elbow, CO
2 cuvette
Critical -
Components Product name Manufacturer Contact time Concentration
Breathing
valve, hoses,
and pressure
sensor
Neodisher LM2 Dr. Weigert 20 minutes 2 %
Flow sensor Sekusept
Powder Classic
Ecolab 15 minutes 4 %
Breathing
valve, hoses,
pressure sen-
sor, and flow
sensor
Korsolex Extra Bode Chemie 30 minutes 3 %
Process step Product name Manufacturer Contact time Temperature
Machine clean-
ing
Neodisher
MediClean
Dr. Weigert At least
5 minutes
40 to 60 °C
(104 to 140 °F)
Machine disin-
fection (ther-
mal)
- - 10 minutes 93 °C (199 °F)
Components Product name Manufacturer Contact time Concentration
Device, O
2
hoses, acces-
sories bag, belt
Dismozon Plus Bode Chemie 15 minutes 1.6 %