Instructions for use | Oxylog VE300 SW 1.n 115
Technical data
15.6 Displayed measured values
The specified accuracy only applies for the specified display range.
Patient port 22 mm ISO conical connector
1) At supply pressure >350 kPa (50.76 psi).
The maximum inspiratory flow is reduced to 80 L/min at supply pressures <350 kPa (50.76 psi) and to 39 L/min at
supply pressures <280 kPa (40.61 psi).
2) When using an accessory with a female connector, add 2 mL to the breathing circuit dead space.
Airway pressure measurement
Display range 0 to 100 mbar (0 to 100 cmH
2O)
Resolution 1 mbar (1 cmH
2O)
Accuracy ±(2 mbar [2 cmH
2O] + 8 % of measured value)
Flow measurement
Minute volume MVe
Display range 0 to 100 L/min, BTPS
Resolution 0.1 L/min
Accuracy ±20 % of measured value or ±0.4 L/min,
whichever greater
Tidal volume VTe
Display range 0 to 5000 mL, BTPS
Resolution 1 mL
Accuracy ±20 % of the measured value or ±20 mL,
whichever greater (adult breathing hose)
CO
2 measurement (option)
Measurement principle Mainstream system
Display range 0 to 100 mmHg / 0 to 13.2 Vol% /
0 to 13.3 kPa
Resolution 1 mmHg / 0.1 Vol% / 0.1 kPa
Accuracy Automatic ambient pressure compensation
No later than 3 minutes after switching on:
Reusable cuvette:
–20 to +40 °C: < ±(0.44 kPa + 8 % relative)
(–4 to +104 °F: < ±[3.3 mmHg + 8 % relative])
+40 to +50 °C: < ±0.60 kPa + 5 % relative)
(+104 to +122 °F: < ±[4.5 mmHg + 5 % relative])