Instructions for use | Oxylog VE300 SW 1.n 51
Getting started
4. Touch the button System test with breathing circuit (B).
The system test with breathing circuit can be canceled at any time by
touching the X button in the top right-hand corner of the screen.
5. On screen 1/4 confirm the advisory messages 1. Make sure that no patient is
connected. and 2. Connect the O
2 supply. with the Next button.
6. On screen 2/4 confirm the advisory messages 3. Connect the appropriate
breathing circuit. and Open the gas outlet. with the Next button.
The system test with breathing circuit is then started. Its progress is indicated by
a bar on screen 3/4.
7. When prompted Seal the patient port., seal the patient port.
The test result is displayed on screen 4/4.
8. To exit, touch the Finish button in the bottom right-hand corner of the screen.
6.3.3 Troubleshooting
WARNING
Risk of patient injury
If the system test or system test with breathing circuit was not completed
successfully, the ventilator is not ready for operation.
Never use a ventilator that is not ready for operation.
If one of the system tests was not completed successfully:
1. For further information, see the following chapter: "Error messages during the
system test", page 97.
2. Contact DrägerService.
6.4 CO2 zero calibration and CO2 filter test (option)
The CO2 zero calibration and the CO2 filter test only work when the CO2 option has
been installed and the CO
2 sensor is present. The sensor temperature must be
between 10 °C and 40 °C (50 °F and 104 °F).
The CO
2 zero calibration and the CO2 filter test can be performed in standby mode
and during ventilation.
The following tests of the CO
2 sensor are recommended:
30184
B
Test Test interval
CO
2 zero calibration Weekly
CO
2 filter test Monthly