116 Instructions for use | Oxylog VE300 SW 1.n
Technical data
15.7 Monitoring
Disposable cuvette:
NOTE: Measurement uncertainty increases at
temperatures <0 °C (+32 °F).
–20 to -10 °C: < ±(0.7 kPa + 22 % relative)
(–4 to +14 °F: < ±[5.3 mmHg + 22 % relative])
–10 to 0 °C: < ±(0.44 kPa + 13 % relative)
(+14 to +3F < ±[3.3mmHg + 13% relative])
0 to +40 °C: < ±(0.44 kPa + 8 % relative)
(+32 to +104 °F: < ±[3.3 mmHg + 8 % relative])
+40 to +50 °C: < ±(0.60 kPa + 5 % relative)
(+104 to +122 °F: < ±[4.5 mmHg + 5 % relative])
Total system response
time
200 ms
Respiratory rate measurement
Display range 0 to 99 /min
Resolution 1 /min
Accuracy ±1 /min
Waveform display
Airway pressure Paw (t) 0 to 90 mbar (0 to 90 cmH
2O)
Flow (t) -150 to 150 L/min
CO
2 0 to +100 mmHg / 0 to +15 Vol% /
0 to +15 kPa
Expiratory minute volume MVe (option)
Alarm, upper alarm limit If the upper alarm limit has been exceeded
Setting range 2 to 60 L/min
Alarm, lower alarm limit If the value has fallen below the lower alarm limit.
Setting range 0.5 to 40 L/min
Alarm delay 40 seconds after the start of ventilation
Airway pressure Paw
Alarm, upper alarm limit If the value Pmax has been exceeded by
+5 mbar (+5 cmH
2O)
Setting range 20 to 60 mbar (20 to 60 cmH
2O)
Alarm, lower alarm limit If the differential pressure between inspiratory
and expiratory phases is too low.
Apnea
Alarm If no breathing phase change is detected for
>15 seconds