56 Instructions for use | Oxylog VE300 SW 1.n
Operation
Step 1: Select the patient category
Step 2: Select the ventilation category
Step 3: Start ventilation
The settings in the ventilation category (step 2) are preset but can be adapted.
For further information, see the following chapter: "Advanced system
setup", page 78.
The default settings for the patient category Child correspond to a child weighing
approx. 25 to 33 kg (55.1 to 72.8 lb) with a tidal volume of 6 to 8 mL per kilogram of
body weight, which corresponds to an age of 7 to 10 years.
The minimum configurable tidal volume of 100 mL corresponds to a child
weighing approx. 13 to 17 kg (28.7 to 37.5 lb) with a tidal volume of 6 to 8 mL per
kilogram of body weight, which corresponds to an age of 2 to 4 years.
7.3 Selecting the ventilation mode
The ventilation mode is selected based on the selected ventilation category:
7.3.1 VC-CMV, VC-AC
VC-CMV
Volume-Controlled - Controlled Mandatory Ventilation
Volume-controlled ventilation with a fixed mandatory minute volume set by the tidal
volume VT and the respiratory rate RR. For patients who are not breathing
spontaneously.
Ventilation category Prerequisite Ventilation mode
CPR None VC-CMV
Ventilation (volume) None VC-CMV/VC-AC
Option is activated and if set
as default
VC-SIMV
Ventilation (pressure) Option PC-BIPAP
CPAP None SPN-CPAP
Option SPN-CPAP/PS
28239
AB
C