Instructions for use | Oxylog VE300 SW 1.n 47
Getting started
6 Getting started
6.1 Charging the battery
The battery temperature must be between 0 and 45 °C (32 °F and 113 °F) when
charging the battery.
WARNING
Risk of patient injury
The battery is not charged outside the specified temperature range.
Observe the permitted temperature when charging the battery. For further
information, see the following chapter: "Operating data", page 118.
The LED next to the battery symbol (A) lights up when an external power source is
connected.
A two-colored indicator (A) lights up to show the current state of battery charge:
Flashing green: The battery is charging.
Lit green: The battery is fully-charged.
Yellow: A battery is not inserted or cannot be charged.
The LED beside the battery symbol remains unlit while the ventilator is being
operated by the battery.
An external battery charger connected to the mains power supply can be used to
charge an extra battery. For further information, see the following chapter: "List of
accessories", page 125.
Display of the battery charge during battery operation
When the device is switched on, the remaining battery charge is displayed in the
header bar below the oxygen mode as a symbol, in increments of 25 percent. In
addition, the remaining duration of operation (H:MM) is displayed (A). The numeric
display appears after a short delay on startup.