Instructions for use | Oxylog VE300 SW 1.n 119
Technical data
DC/DC converter
Gas supply
15.9 Materials used
Housing
Reusable breathing circuit
The power supply unit is intended for indoor use
only (e.g., in hospitals or fire stations).
Fuses F1 and F2 T2.5 AH / 250 V~
Protection class (as defined in
IEC 60601-1)
Class II
Degree of protection IP 42
Input 12 / 24 / 28 VDC; 5 A / 2.5 A / 2.1 A
Output 19 V / 2.6 A
The intended use of the DC/DC converter is in
vehicles.
Fuse Type: FP1 MINI Style PCB
Voltage: 32 VDC / 10 A
Breaking capacity: 1000 A / 32 VDC
Operating speed: fast-acting
From a central gas supply system or an oxygen
cylinder
Oxygen supply pressure 270 to 690 kPa (39.16 to 100.08 psi)
Supply gas Medical oxygen, oxygen 93 %
Connection to the oxygen
supply
either:
NIST (Non-Interchangeable Screw-Threaded) as
per EN 739 / ISO 5359 or
DISS (Diameter Index Safety Systems) as per
CGA V5-1989 or
NF (Norme franҫaise) S90-116
Specific quick-coupling port
Device Acrylonitrile styrene acrylate/polycarbonate
(ASA/PC), thermoplastic copolyester elastomer
(TPC)
Base plate of the carrying
system
Polyamide (PA)
Power supply unit Polycarbonate (PC)
DC/DC converter Polycarbonate (PC)
Breathing hose Silicone rubber
Housing of flow sensor Polysulfone (PSU)