86 Instructions for use | Oxylog VE300 SW 1.n
Logbook and system information
10 Logbook and system information
10.1 Logbook
To access the logbook:
Standby mode
1. Touch the button 1 in the top right-hand corner of the screen to go to screen 2.
2. Touch the button Logbook (A).
Operation mode
1. To access the start dialog, touch the X button in the top right-hand corner of the
ventilation dialog.
Alternatively: Press the start dialog key on the control panel.
2. Then touch the button 1 in the top right-hand corner of the screen to go to
screen 2.
3. Touch the button Logbook (A).
The logbook displays all events, alarms, and the switching off of the device in
chronological order. The most recent entry is at the top. Earlier events and alarms
can be displayed using the arrow keys.
A maximum of 400 logbook entries is visible. When the capacity of the logbook is
reached, the oldest entries are overwritten. For alarms, the presence of an alarm
condition and the removal of this condition is recorded. Alarms are displayed in the
color corresponding to their priority and the removal of the alarm condition is
displayed in gray.
The entries in the logbook cannot be deleted and are preserved even after the
device is switched off and on or after a power supply failure.
2939633072