EasyManua.ls Logo

Dräger Oxylog VE300 - Device Operation; Switching Device On and Off

Dräger Oxylog VE300
134 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Instructions for use | Oxylog VE300 SW 1.n 55
Operation
7 Operation
7.1 Switching on the device
WARNING
Risk of patient injury and device malfunction
Possible health risks for patients and users.
Use only cleaned and disinfected ventilators that have been successfully
checked as being ready for operation. For further information, see the following
chapter: "Reprocessing", page 98.
Do not cover the device (e.g., with a blanket).
To switch the device on:
1. Briefly press the start/standby key (A).
The device briefly displays the startup screen, indicating the device name and the
installed software version.
Automatic self-test
Every time it is switched on, the device performs an automatic self-test involving a
test of the acoustic alarm signals as well as an internal device test. The device then
switches to standby mode.
7.2 Preparing the ventilation mode
The desired ventilation mode is activated by 3 steps from standby mode:
Fig. 1 Selection in 3 steps
28576
A
28210

Table of Contents

Other manuals for Dräger Oxylog VE300

Related product manuals