104 Instructions for use | Oxylog VE300 SW 1.n
Reprocessing
12.6 Reprocessing the non-critical components
12.6.1 Surface disinfection with cleaning
WARNING
Risk due to penetrating liquid
Penetrating liquid may cause the following:
Damage to the device
Electric shock
Device malfunctions
Ensure that no liquid penetrates the device.
1. Remove soiling immediately. Use a cloth dampened with disinfectant to remove
soiling.
2. Perform surface disinfection.
3. After the product has been exposed to the disinfectant for the specified contact
time, remove residual disinfectant.
4. Wipe with a cloth dampened with water (preferably drinking-water quality). Allow
the product to dry.
5. Check the product for visible soiling. Repeat steps 1 to 5 if necessary.
6. Check the product for visible damage and replace if necessary.
12.7 Reprocessing list
The instructions of the hospital’s infection control officer shall prevail and must be
observed by the user.
Components Surface
disinfec-
tion with
cleaning
Manual
cleaning
followed by
disinfec-
tion by
immersion
Machine
cleaning
with ther-
mal disin-
fection
Steam
steriliza-
tion
Recom-
mended
reprocess-
ing intervals
Device and
O
2 hoses
Yes No No No After every
patient / if
soiled
Reusable
breathing
circuit
No Yes Yes Yes After every
patient / if
soiled
CO
2 sensor Yes No No No After every
patient / if
soiled
Reusable
cuvette of the
CO
2 sensor
No Yes Yes Yes After every
patient / if
soiled
Test filter for
CO
2 sensor
Yes No No No After every
patient / if
soiled