Instructions for use | Oxylog VE300 SW 1.n 113
Technical data
Battery
Compressed gas hose (Alduk pressure reducer to device port)
15.4 Settings
Using the therapy controls the values can be adjusted without loss of accuracy. The
controlled parameters pressure, flow and volume can only be applied with the
accuracy of the corresponding measured value.
Temperature –20 to +50 °C (–4 to +122 °F)
(preferred long-term storage temperature <35 °C
[<95 °F])
Ambient pressure 620 to 1100 hPa (8.99 to 15.95 psi)
Relative humidity 5 to 95 % (without condensation)
Temperature –30 to +55 °C (–22 to +131 °F)
Ambient pressure 620 to 1100 hPa (8.99 to 15.95 psi)
Relative humidity 5 to 95 % (without condensation)
Ventilation modes VC-CMV, VC-AC, SPN-CPAP
Optional: VC-SIMV/PS, PC-BIPAP/PS,
SPN-CPAP/PS
Respiratory rate RR
Adult 2 to 30 /min (VC-SIMV, PC-BIPAP)
5 to 30 /min (VC-CMV, VC-AC)
Child 2 to 50 /min (VC-SIMV, PC-BIPAP)
10 to 50 /min (VC-CMV, VC-AC)
Maximum airway pressure Pmax 20 to 60 mbar (20 to 60 cmH
2O)
Ratio of inspiratory time to expiratory time I:E 1:4 to 4:1
Inspiratory time Ti 0.3 to 10 s
Tidal volume VT
Adult 0.3 to 2.0 L, BTPS
Child 0.1 to 0.4 L, BTPS
Measurements referred to conditions of the
patient's lungs, body temperature 37 °C (98.6 °F),
ambient pressure, water vapor saturated gas.
Accuracy ±20 % of the setting
O
2 concentration 100 % O2 and O2/air mix
The actual value depends on the inspiratory flow
and the mean airway pressure.
Inspiratory pressure Pinsp PEEP +3 to +55 mbar
Positive end-expiratory pressure PEEP 0 to 20 mbar (0 to 20 cmH
2O)
Trigger sensitivity
(flow trigger)
Off or 3 to 15 L/min