Instructions for use | Oxylog VE300 SW 1.n 97
Troubleshooting
11.4 Error messages during the system test
Message Cause Explanation/Remedy
Test failed during gas
outlet test.
The gas outlet was not fully
opened.
Fully open the gas outlet
when prompted during the
system test.
Contact DrägerService.
Test failed during
leakage test.
Leakage has occurred. Seal the gas outlet more
tightly when prompted
during the system test.
Contact DrägerService.
Test failed during
patient port test.
The patient port was not
fully opened.
Fully open the patient port
when prompted during the
system test.
Contact DrägerService.
Test failed. The sensor reports a CO
2
filter test error.
Perform the zero calibra-
tion.
Do not breathe onto the
CO
2 sensor during the zero
calibration.
Clean the CO
2 test filter and
the CO
2 sensor.
Repeat the filter test.
A malfunction of the pneu-
matic components has
occurred.
Contact DrägerService.
Test failed. Check
breathing circuit
type.
The detected hose type
does not match the selected
hose type.
Select a different hose type.
Connect a different hose
type.
Test failed. Check
breathing circuit.
An error occurred during the
breathing circuit test.
Check the measuring lines.
Check the breathing circuit
and the connections.
Replace the breathing cir-
cuit.
Zero calibration
failed.
The CO
2 sensor has a hard-
ware failure.
Replace the CO2 sensor.
The sensor windows are
soiled.
Clean the CO
2 sensor win-
dows.
An error during the zero cal-
ibration of the CO
2 sensor
occurred.
Repeat the zero calibration.
Do not breathe onto the
CO
2 sensor during the zero
calibration.