Instructions for use | Oxylog VE300 SW 1.n 35
Assembly and preparation
5. Connect the oxygen compressed gas hose to the pressure reducer and the
ventilator.
6. Rotate the cylinder so that the pressure reducer and the hose to not protrude
beyond the bar.
7. Attach and fasten both closures until they engage audibly.
WARNING
Risk of patient injury
If the oxygen cylinder is not properly mounted, it may slip out of the carrying
system.
► Make sure that the oxygen cylinder is mounted securely in the carrying
system.
WARNING
Risk of patient injury and damage to the device
After the device falls with the carrying system, the closures, hinges and bar may no
longer be in proper functioning order to safely hold the oxygen cylinder.
► Check the closures, hinges, and protection bar to ensure functional integrity and
replace them if necessary.
5.2 Attaching the accessories bag (option)
1. Pull on the buttons fastened inside the accessories bag.
2. Push the outside openings of the buttons onto the pins on the side of the device.
3. Release the buttons.
5.3 Internal battery
Internal power is provided by means of a removable rechargeable battery. For
further information, see the following chapter: "Operating data", page 118.
NOTICE
► To activate: Fully charge the battery before commissioning.
Inserting the battery
1. Remove the battery compartment cover.
2. Insert the battery vertically in the battery compartment with the contacts facing
downward until the locking lever on the carrying handle side engages.
3. Fit the battery compartment cover and press it until it engages audible.
Removing the battery
1. Remove the battery compartment cover.
2. Press the locking lever on the carrying handle side.
3. Lift out the battery vertically using the strap.