52 Instructions for use | Oxylog VE300 SW 1.n
Getting started
6.4.1 Performing the CO2 zero calibration (option)
Do not breathe on the CO2 sensor during zero calibration, otherwise the
procedure may fail or may be performed with an invalid zero value.
1. To switch the ventilator on, briefly press the start/standby key.
The ventilator performs an automatic self-test, then the standby screen is
displayed.
2. Connect the CO
2 sensor to the device.
3. To move to screen 2 of the dialog, touch the button 1 in the top right-hand corner
of the screen.
4. Touch the button Test & calibration (A).
5. Touch the button CO
2 zero calibration (B).
The CO2 zero calibration can be canceled at any time by touching the
X button in the top right-hand corner of the screen.
Shortly after connecting the CO2 sensor, the zero calibration may take longer
(up to 2 minutes).
6. Remove the CO
2 sensor (A) from the CO2 cuvette (B) of the breathing circuit.