Instructions for use | Oxylog VE300 SW 1.n 39
Assembly and preparation
Connecting the device to a central gas supply
1. Connect the oxygen compressed gas hose (A) to the standard port (B) or to the
additional quick-coupling port of the device.
2. Connect the gas probe (C) to the oxygen terminal unit until it locks in place and
the oxygen supply is assured.
Connecting the built-in oxygen cylinder
For information on the correct assembly of the main device with carrying system,
see page 33.
WARNING
Risk of ventilator malfunction
Components installed in the gas supply line might cause malfunctions.
Do not install dosage valves or flowmeters in the device's gas supply line.
To avoid insufficient oxygen supply during use, always check the cylinder's
oxygen pressure before use.
30242
B
A
C