Operation
18
Operation
Run equipment check according to checklist, see page 96.
Starting Up
Switching on
1 Turn the main switch on the back of the apparatus to » «
The Savina now runs through a power-up self-test
procedure.
Wait for the 12-second test phase to be completed.
At the end of the test phase, the apparatus automatically
starts ventilation with the last ventilation mode set and the
associated ventilation parameters.
During the self-test the start-up page appears, showing the
software version.
After this, the main page appears.
2 Check the settings in the display fields next to the
parameter keys and adjust as necessary.
To adjust ventilation parameters:
Either
2 Change the ventilation parameters after the Savina has
started ventilation
or
3 Press the » Standby« switch to set the Savina on
standby,
4 Confirm = press »Alarm Reset« key.
2 Change the ventilation parameters and
3 press the » Standby« switch to return to ventilation.