Care
71
Disinfecting/Cleaning/Sterilising
Use surface disinfectants. For surface compatibility, use
disinfectants based on:
— aldehydes,
— quaternary ammonium compounds.
Due to possible damage to the materials, do not use any
preparations based on:
— Compounds containing alkylamine
— Compounds containing phenol
— Halogen-releasing compounds
— Strong organic acids
— Oxygen-releasing compounds
For users in the Federal Republic of Germany, we recommend
that only disinfectants on the current DGHM list are used
(DGHM: German Society for Hygiene and Microbiology).
The DGHM list (published by mhp-Verlag, Wiesbaden) also
classifies each disinfectant by its active agents. For countries
where the DGHM list is not available, we recommend the types
of disinfectant given above.
Disinfectants often contain – besides their main active agents
– additives that can also damage materials.
● If in doubt, ask the supplier/manufacturer of the
disinfectant/cleaning agent.
— Savina basic device with trolley
— Hinged arm
— Medical gas hose
● Wipe disinfect.
— Ventilation hoses
— Y-piece
— Water traps and collecting jars
— Expiration valve casing
— Expiration valve diaphragm
— Containers for water traps
● Hot-steam disinfect in the autoclave (93
o
C/10 minutes)
using detergent only.
● Do not sterilise parts in ethylene oxide!
Ethylene oxide can diffuse into the parts.
Danger to health!