Operation
32
Advisory
Low priority message
1 Yellow LED lights up and remains lit.
Advisory messages are marked with one exclamation mark.
Example: ! Flow monitoring off
Savina generates a two-tone sequence that sounds only once.
To remedy the faults, please refer to the "Troubleshooting"
section starting on page 56.
When the fault has been remedied
— the audible alarm is switched off.
— Caution messages (!!) and advisory messages (!) disappear
automatically from the screen.
Warning messages (!!!) remain black on a white background on
the screen and need to be acknowledged:
2 Press the »Alarm Reset« key.
3 The message is erased from the screen.
Suppressing the audible alarm
for max. 2 minutes:
4 Press the » « key. Its yellow indicator LED lights up and
the alarm tone will be muted for about 2 minutes. If the
cause of the alarm has not yet been remedied, the audible
alarm will sound again after about 2 minutes.
If you wish to reactivate the audible alarm before the end of this
2-minute muting period:
4 Press the » « key again. The yellow LED will now go out.
The message remains on the screen.
The alarm tone sounds only once when a higher priority alarm
occurs during the period, when the audible alarm is muted.