37
Special functions
Manual inspiration
This function may be used in all modes except CPAP
spontaneous breathing without ASB pressure support.
Regardless of the start time, an automatic ventilation stroke
can be prolonged by up to a maximum of 15 seconds.
Or:
Between two automatic ventilation strokes, a ventilation
stroke can be manually started and held for a maximum
of 15 seconds.
The pattern of the manually started ventilation stroke depends
on the ventilation mode in use.
For IPPV, SIMV:
volume-controlled ventilation stroke, defined by the »VT« and
»Tinsp« settings.
For BIPAP:
pressure-controlled ventilation stroke, defined by the »Pinsp«
and »Tinsp« settings.
For CPAP/ASB:
pressure-assisted ventilation stroke, defined by the
»PASB above PEEP« setting.
1 Press and hold down the »Insp. hold« key for as long as
inspiration is required.
Either an automatic ventilation stroke that has just begun will be
accordingly prolonged, or a new ventilation stroke can be
started and held for a maximum of 15 seconds.