Configuration
48
Configuration
The following settings for use can be selected in
»Configuration« mode:
— Screen contrast
— Alarm volume
— Measured values line
— Manual calibration for O
2 sensor 2
— FiO2 and flow monitoring ON/OFF
— Pmax ON/OFF
— Plateau ON/OFF
— Language, date and time
— MEDIBUS protocol.
The settings selected in »Configuration« mode remain stored
until the equipment is switched off.
The following information about the equipment can be
displayed in »Configuration«:
— Total operating hours
— Operating hours since the last inspection and
maintenance.
— software version
— device identification number
Setting the screen contrast
● Press »Config. ∆∆« key repeatedly until
»Configuration 1/4« appears.
● Select the »Contrast« line = turn the rotary knob,
activate = press the rotary knob
● To set the contrast = turn the rotary knob.
The contrast changes.
Confirm = press the rotary knob.
The contrast set will now be activated.
Adjusting the volume of the audible alarm
Display (example):
● Press »Config. ∆∆« key repeatedly until
»Configuration 1/4« appears.
● Select »Volume« line = turn the rotary knob,
activate = press the rotary knob.
● To adjust the volume = turn the rotary knob,
confirm = press the rotary knob
After the setting has been confirmed, the alarm tone
is sounded once to enable you to judge the volume.
077 37388122 37388