Operation
34
In »Mask/NIV« mode, another page can be displayed
»Values 3/3«.
NIV – Non-invasive ventilation
Mask ventilation (optional)
Installing the NIV option
This should only be done by experts, following the
corresponding installation instructions.
Using NIV
*
All ventilation modes are selectable in »Mask/NIV« mode.
Selecting »Mask/NIV« mode
Switch apparatus to standby = hold down » Standby«
key for about 3 seconds.
Press »Alarm Reset« to switch off the audible alarm.
Select »Patient Connection« line = press rotary knob,
Select »Mask/NIV« line = press rotary knob,
Confirm = press rotary knob.
When masks are used, the dead space increases.
Follow the mask manufacturer’s directions.
* For a detailed description of NIV, see page 106.
Never ventilate an intubated patient in the »Mask/NIV«
mode.
After changing over from »Mask/NIV« to »Tube« mode,
check the alarm limits.
Avoid high airway pressures.
Risk of aspiration.