Care
74
Maintenance Intervals
Clean and disinfect equipment and/or components
before any maintenance procedures – and before returning
for repair!
O2 sensors replace when the following message is displayed:
!!! O2 measurement inop.
in case of calibration is no longer possible.
Used O2 sensors can be returned to
Dräger Medical AG & Co. KGaA for disposal.
Microfilter Replace after 1 year, see page 75.
Dispose of with normal domestic waste.
Dust filter set Replace after 1 year, see page 76.
Dispose of with normal domestic waste.
Blower unit To be replaced by trained service personnel after
20.000 hours’ operation or after 8 years – whichever is earlier.
Lead-gel battery To be replaced by trained service personnel after 2 years.
Lead-gel batteries should be disposed of as special waste,
see page 78.
Filter in O
2 inlet To be replaced by trained service personnel every 6 years.
Real-time clock To be replaced by trained service personnel every 6 years.
Pressure reducer To be replaced by trained service personnel every 6 years.
Equipment inspection
and service
Should be carried out for the first time after 2 years or
6,000 hours’ operation, whichever is earlier.
Thereafter annually or after 6,000 hours’ operation, whichever
is earlier.
For other servicing (e.g. six-monthly inspection, etc.), contact
DrägerService.
Only original Dräger parts should be used.